1 / 5
| Model | PA-900B (Advanced Model) |
| Display | 10.4 inch high visibility color TFT display |
| Ventilation Mode | A/C, SIPPV, IPPV, SIMV, PSIMV, PSV, CPAP, PCV, VCV, SPONT MANUAL |
| Minute Volume | ≥ 18L |
| Back-up Power Supply | At least 4 hours |
| Tidal Volume | Adjustable: 20-1500ml, Display: 0-2000ml |
| Respiratory Rate | 1-99bpm |
| I:E Ratio | 4:1 - 1:4 |
| PEEP | 0-10cmH₂O |
| SIMV Rate | 1-20bpm |
| Inspiratory Pressure Trigger | -10 - 10cmH₂O |
| Trigger Sensitivity (flow) | 1-10L/m |
| Inspiratory Plateau | Adjustable from 0-50% of inspiratory time |
| Adjustable Oxygen Concentration | 21-100% |
| Gas Supply Requirement | 280~600 kPa medical grade oxygen and compressed air |
| Maximum Safe Operating Pressure | ≤ 6.0kPa |
| Tidal Volume Alarm | Upper: 10~2000ml, Lower: 0~1800ml |
| Airway Pressure Alarm | Upper: 0.1~6kPa, Lower: 0~5kPa |
| Oxygen Concentration Alarm | Upper: 22~100%, Lower: 21~80% |